Hioki MR8870 Memory HiCorder

Sale price0,00 kr Price on quotation
Available

Use our chat for personal support. Or contact us via +45 31 33 18 19 or salg@GOmeasure.dk


Hioki Memory HiCorders are high-speed waveform monitoring and recording devices that can simultaneously measure multiple quick-changing phenomena due to complete isolation between channels and from each channel to the ground. The MR8870 is a handheld data acquisition (DAQ) recorder built in with 2 analog channels and 2MW of memory, suitable for elevator maintenance, machine, motor and inverter fault analysis, testing and inspections.

Discover the possibilities

More information

Hioki MR8870 Memory HiCorder

Hioki Memory HiCorders are high-speed waveform monitoring and recording devices that can simultaneously measure multiple quick-changing phenomena due to complete isolation between channels and from each channel to the ground. The MR8870 is a handheld data acquisition (DAQ) recorder built in with 2 analog channels and 2MW of memory, suitable for elevator maintenance, machine, motor and inverter fault analysis, testing and inspections.

 

Key Features

• Mode for recording instantaneous waveform and RMS fluctuations

• Save values in real time to a CF card

• Record four channels at once by synchronizing two instruments with the bundled PC application

• Compact and easy to carry

• Easy, intuitive operation

• Fast, 1MS/s performance despite the compact size

• Built-in, compact-yet-sharp QVGA-TFT wide LCD

 

■ Basic specifications (Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)

Number of channels2 analog channels + 4 logic channels (standard)\nNote: Isolated analog channels, isolated input and frame, logic has common GND
Measurement ranges10 mV to 50 V/div (10 div full-scale), 12 ranges, Resolution: 1/100 of range
Max. rated voltageBetween terminals: 400 VDC, Between terminal to earth: 300 VAC, DC CAT II
Frequency characteristicsDC to 50 kHz (-3 dB)
Time axis (Memory mode)100 µs to 5 min/div, 20 ranges,at 100 points/div resolution, three steps of time-axis magnification from ×2 to ×10, and 9 steps of time-axis compression from ×1/2 to ×1/1,000
Recording intervals (RMS mode)1 ms to 1 min., 16 settings, sampling period: 200 μs (fixed) (for AC voltage/current, 1,000 RMS values/sec.), envelope mode always on\nNote: Only the maximum value and minimum value for each recording interval are recorded.
Measurement functionsMemory recorder (high speed recording), RMS recorder (50/60 Hz, DC only)
Memory capacity12-bits × 2M-words/ch (1 word = 2 bytes)
Removable storageCF card TYPE I slot ×1 (Up to 2 GB)
Display4.3-inch WQVGA-TFT color LCD (480 × 272 dots)
Displayable languagesEnglish, Japanese
InterfacesUSB 2.0 mini-B receptacle ×1, Functionality: Connect the instrument to a PC to send files on the CF card to the PC. The instrument cannot be controlled from a PC.
PrinterN/A
Power supplyAC Adapter Z1005: 100 to 240 VAC (50/60 Hz) , 30 VA max. (when using the AC adapter and charging the 9780 with the instrument)\nBattery Pack 9780: 3 VA, continuous operating time of approx. 2 hr. (25°C reference value; when used with the Z1005, the Z1005 takes priority), charging time of 200 min. using the AC adapter (25°C reference value) (option)\nExternal DC power: 10 to 16 V, 10 VA max. (connection cord of 3 m or less is available by special-order)
Dimensions and mass176 mm (6.93 in)W × 101 mm (3.98 in)H × 41 mm (1.61 in)D, 600 g (21.2 oz) (with the Battery pack 9780 installed)
AccessoriesInstruction manual ×1, Measurement guide ×1, AC adapter Z1005 ×1, Strap ×1, USB cable ×1, Application disk (Dedicated program for the MR8870) ×1, Protection sheet 9809 ×1